Quantcast
Channel: Metabolic Rate Test – Microlife MedGem and BodyGem Indirect Calorimeters
Viewing all articles
Browse latest Browse all 39

A Comparison of the BodyGem and the VacuMed VO2 Lab for Measuring Resting Metabolic Rate.

$
0
0

American College of Sports Medicine, Regional (Central States Chapter) Conference

Project: Accuracy of Indirect Calorimetry and Predictive Equations for Measuring Resting Metabolic Rate. October 2016

Authors:

  • Lucas Kyle Shrum – Southeast Missouri State University
  • Karl Keczkemethy – Western Kentucky University
  • Thomas Pujol – Southeast Missouri State University
  • Jason D. Wagganer – Southeast Missouri State University

Abstract

Indirect calorimetry is used to measure oxygen consumption as a means of accurately estimating resting metabolic rate (RMR). Many commercial devices are available for measuring RMR, such as the BodyGem® indirect calorimeter and the VacuMed VO2 Lab metabolic measuring system.

Metabolic carts, such as those manufactured by VacuMed and CosMed, tend to be viewed as the criterion method for cardiopulmonary exercise testing equipment.

However, newer, less expensive, and more portable devices, like the BodyGem® and MedGem®, are being used in non-laboratory settings to measure RMR.

PURPOSE: The purpose of this study was to compare the measured RMR of college-aged participants using a BodyGem® indirect calorimeter and a VacuMed VO2 Lab metabolic cart.

METHODS: Each participant (male = 15; female = 15; age=22.7 ± 3.3 yrs; wt = 77.5 ± 14.5 kg; ht = 173.5 ± 9.6 cm) completed a thirty-minute supine resting session in a quiet environment. Five minutes of resting data were then collected using the VacuMed VO2 Lab while in the supine position. Immediately thereafter, five minutes of resting data was collected using the BodyGem® indirect calorimeter while in the seated position.

RESULTS: A paired samples t-test revealed a significant difference between the BodyGem® (1995.0 ± 540.5 kcals) and the VacuMed (1520.9 ± 452.5 kcals) (p < 0.001). The BodyGem® was significantly higher for both male (p = 0.001) and female (p = 0.042) participants.

CONCLUSION: Inconsistent with past research, the BodyGem® indirect calorimeter yielded RMR values significantly greater than the VacuMed VO2 Lab metabolic measuring system.

Further research needs to be conducted comparing the various hand-held indirect calorimeter and metabolic cart instruments before methods such as the BodyGem® can be deemed accurate for measuring RMR.

To measure your clients Resting Metabolic Rate, you need an indirect calorimeter.

Microlife MedGem Miramont Family Medicine Case StudyThe Microlife BodyGem and MedGem indirect calorimeters measure Resting Metabolic Rate, to help Bariatric Doctors, Nutritionists, Dietitians and Personal Trainers; optimize the weight management results of their clients.

Microlife Medical Home Solutions is assisting physicians’ and their medical practices with implementing a patient-centered, physician-guided approach to an improved evaluation, diagnosis and treatment of cardiovascular, pulmonary, and metabolic diseases.

The MedGem and BodyGem Indirect Calorimeter by Microlife is a handheld, portable indirect calorimeter that measures oxygen consumption (VO2) and RMR.

Click on the links to get more information on the Microlife BodyGem and MedGem indirect calorimeters to start measuring your clients Resting Metabolic Rate.


Viewing all articles
Browse latest Browse all 39

Latest Images

Trending Articles





Latest Images